5.1.1 Transfer Surface Height Adjustability. The transfer surface should have an adjustable height in small virtually continuous increments.
Search Results "Patient Lift"
-
5.1.1 Transfer Surface Height Adjustability
-
1009.1 General
Where provided, pool lifts, sloped entries, transfer walls, transfer systems, and pool stairs shall comply with 1009....
-
1009.1 General
Where provided, pool lifts, sloped entries, transfer walls, transfer systems, and pool stairs shall comply with 1009....
-
M303.2.1 Orientation
Wheelchair spaces shall be designed so that a patient seated in a wheelchair orients in the same direction that a patient not seated in a wheelchair orients when the diagnostic equipment...
-
M303.2.1 Orientation
Wheelchair spaces shall be designed so that a patient seated in a wheelchair orients in the same direction that a patient not seated in a wheelchair orients when the diagnostic equipment...
-
5.3.3.1 Transfer Sides Recommendation for Exam Chairs with Fixed Footrests
The current design allows only one long side for transfer, which limits some patient transfers where a patient can use only one side of the body due to paralysis on one side or other such...
-
206.7.2 Wheelchair Spaces
Platform lifts shall be permitted to provide an accessible route to comply with the wheelchair space dispersion and line-of-sight requirements of 221 and 802....
-
F206.7.2 Wheelchair Spaces
Platform lifts shall be permitted to provide an accessible route to comply with the wheelchair space dispersion and line-of-sight requirements of F221 and 802....
-
11B-1009.1 General
Where provided, pool lifts, sloped entries, transfer walls, transfer systems, and pool stairs shall comply with Section 11B-1009....
-
1009.1 General
Where provided, pool lifts, sloped entries, transfer walls, transfer systems, and pool stairs shall comply with 1009....
-
11B-1009.1 General
Where provided, pool lifts, sloped entries, transfer walls, transfer systems, and pool stairs shall comply with Section 11B-1009....
-
11B-1009.1 General
Where provided, pool lifts, sloped entries, transfer walls, transfer systems, and pool stairs shall comply with Section 11B-1009....
-
M303.4.1 Breast Platform Adjustability
Breast platforms shall be continuously adjustable from a low height of 26 inches (660 mm) to a high height of 42 inches (1065 mm) above the floor.
-
M303.4.1 Breast Platform Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a mammography breast platform height range of 30 inches high minimum and 42 inches high maximum above the floor. The Access Board received three comments on...
-
More Information on Emergency Planning
What? Information you can use to strengthen your emergency planning and to protect: Your family, Your health, Your safety, Your independence,...
-
Emergency Health Information
*** View and download a blank Emergency Health Information sheet here. *** It’s good practice to carry emergency health information containing your critical health information and...
-
Sample Emergency Health Information
*** View and download the Sample Emergency Health Information sheet here. ***
-
6.3(3)
(3) Each bedroom shall have adequate space to provide a minimum clear floor space of 36 in (915 mm) along each side of the bed and to provide an accessible route complying with 4.3.3...
-
Attitudes
# Attitudes Yes No Unsure Comments and follow up A. We identify our biases regarding participants with disabilities so that we can...
-
6.3(2)
(2) Each one-bed room shall have a minimum clear floor space of 36 in (915 mm) along each side of the bed, and 42 in (1065 mm) between the foot of the bed and the wall.
-
6.3(3)
(3) Each two-bed room shall have a minimum clear floor space of 42 in (1065 mm), preferably 48 in (1220 mm), between the foot of the bed and the wall; 36 in (915 mm) between the side...
-
M305.1 General
Supports shall comply with M305.
-
M305.1 General (Section-by-Section Analysis)
This is an introductory section.
-
M305.1 General
Supports shall comply with M305, as applicable.