Transfer supports shall comply with M305.2.
Search Results "Fixed Equipment"
-
M305.2 Transfer Supports
-
B. M301 Diagnostic Equipment Used by Patients in a Supine, Prone, or Side-lying Position and M302 Diagnostic Equipment Used by Patients in a Seated Position
In the final rule M301 and M302 provide the technical requirements for diagnostic equipment used in the supine, prone, or side-lying position, and diagnostic equipment used by patients in...
-
F226.1 General
surfaces include, but are not limited to, laboratory and work benches, fume hoods, reception counters, teller windows, study carrels, commercial kitchen counters, writing surfaces, and fixed...
-
M303.3.1 Vertical (Section-by-Section Analysis)
The MDE NPRM proposed that for equipment with a change in level at the entry to the wheelchair space, level changes of up to ¼ inch high are permitted to be vertical....
-
DOT’s ADA Regulations
transportation, as well as to any private entity that provides public transportation or that is not primarily engaged in the business of transporting people but operates a demand responsive or fixed...
-
11B-805.4.3 Turning space
A minimum of a 36-inch wide access shall be provided to and along each side of the bed, exam table, procedure table, gurney, lounge chair and/or equipment/apparatus....
-
11B-805.4.3 Turning space
A minimum of a 36-inch wide access shall be provided to and along each side of the bed, exam table, procedure table, gurney, lounge chair and/or equipment/apparatus....
-
M302.4 Lift Compatibility
Diagnostic equipment shall be usable with portable patient lifts and, when in use with such lifts, shall comply with M302.4.1 or M302.4.2. ...
-
Exceptions Considered for Wheelchair Spaces on Raised Platforms
Patients who use scooters may have other options for using equipment with wheelchair spaces on raised platforms....
- seca 664 - electronic wheelchair scale_without comments - seca1840
- CPF Challenge: 7 Minute Workout OFFICIAL TRAILER (2017)
-
M303.3.2 Beveled (Section-by-Section Analysis)
The MDE NPRM proposed that for equipment with a change in level at the entry to the wheelchair space, level changes greater than ¼ inch but not greater than ½ inch would be required to be...
-
M302.4 Lift Compatibility (Section-by-Section Analysis)
(Significant Changes – Lift Compatibility Exception), the Access Board has added an exception for diagnostic equipment that meets the following three criteria: fixed overhead patient lifts...
-
M305.2.2.1 End Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.1 and M305.2.2.1 shall be 15 inches (380 mm) long minimum. Transfer supports shall be positioned along 13 ½ inches...
-
M305.2.4.1 Circular Cross Section
Transfer supports with circular cross sections shall have an outside diameter of 1 ¼ inches (32 mm) minimum and 2 inches (51 mm) maximum.
-
M305.2.1.2 Side Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.2 shall be capable of supporting transfer on each side of the transfer surface.
-
M305.2.1.1 End Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.1 and M302.2 shall be located on the short side (length) opposite the transfer side.
-
M305.2.2.2 Side Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.2 shall be 28 inches (710 mm) long minimum and shall be positioned along the width of transfer...
-
M303.4 Components (Section-by-Section Analysis)
The MDE NPRM proposed to require diagnostic equipment used by patients seated in a wheelchair which has components that are used to examine specific body parts, be capable of examining those...
-
Knee & Toe Space [M303.2.4]
Each of these measurements creates a more accessible piece of imaging equipment balanced with the physical attributes and operation of the equipment....
-
M303.2.3.1 Front or Rear Entry
Where wheelchair space entry and exit is provided at only one end (front or rear) the wheelchair space shall be 48 inches (1220 mm) deep minimum.
-
M303.2.3.3 Side Entry
Where wheelchair space entry is only from the side, the wheelchair space shall be 60 inches (1525 mm) deep minimum.
-
M303.2.3.2 Pass Through Entry
Where wheelchair space entry and exit permits pass through from one end to the other, the wheelchair space shall be 40 inches deep (1015 mm) minimum.
-
§ 1194.23(i)
(i) Interference to hearing technologies (including hearing aids, cochlear implants, and assistive listening devices) shall be reduced to the lowest possible level that allows a user...