5.5.1 Armrest Provision for Equipment Used by Patient in Seated Position. Armrests are not required but if provided cannot obstruct transfer supports.
Search Results "Examination"
-
5.5.1 Armrest Provision for Equipment Used by Patient in Seated Position
-
5.3.3 Transfer Sides Recommendations for M302
This provision will coincide with the seat area design of most examination chairs. NPRM Proposed Provision: M302.2.3 Transfer Sides....
-
M302.2.3 Size
Transfer surfaces shall be 21 inches (610 mm) wide minimum and 17 inches (430 mm) deep minimum. The size of transfer surfaces shall be measured from center points of their opposing...
-
M302.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and comply with M305.4.
-
M302.3 Supports
Transfer supports, leg supports and reclining surfaces shall comply with M302.3.
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2.
-
M302.3.1 Transfer Supports
Transfer supports shall be provided for use with transfer sides required by M302.2.4 and shall comply with M305.2.1.1, M305.2.2.1, and M305.2.3 through M305.2.8.
-
M307 Operable Parts
[See subsections ...]
-
M302.3.1 Transfer Supports (Section-by-Section Analysis)
In the MDE NPRM the Access Board proposed that transfer supports must be provided for use with transfer sides on diagnostic equipment used by patients in the seated position, and...
-
M302.2.3 Size (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface size for diagnostic equipment used by patients in the seated position of 21 inches wide and 15 inches deep (proposed M302.2.2). The Access...
-
M302.2 Transfer Surface (Section-by-Section Analysis)
This is an introductory section.
-
M302.3 Supports (Section-by-Section Analysis)
This is an introductory section. An editorial change was made to this section as a result of the change in M302.3.2, described below, to replace the word “stirrups” with the term...
-
5.2.2.2 Transfer Surface Depth Recommendation for M302
5.2.2.2 Transfer Surface Depth Recommendation for M302 The Committee recommends a minimum depth requirement of 17 inches for equipment used by patients in seated...
-
5.2.2 Transfer Surface Size for Equipment Used by Patients in Seated Position
5.2.2 Transfer Surface Size for Equipment Used by Patients in Seated Position.
-
5.2.2.2 Transfer Surface Depth
5.2.2.2 Transfer Surface Depth. The transfer surface should be 17 inches deep minimum for equipment used by patients in a seated position.
-
M302.2.2 Size
The transfer surface shall be 21 inches (610 mm) wide minimum and 15 inches (381 mm) deep minimum. Advisory M302.2.2 Size. The size requirements in this section apply only to the...
-
M302.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M302.2.3 and shall comply with M305.2.
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2.
-
M302.3 Supports
Transfer supports, armrests, and reclining surfaces shall comply with M302.3.
-
M307 Operable Parts
[See subsections ...]
-
4. Exception from the Requirements of M301 for Certain Examination Chairs that Comply with M302
The Access Board concurs with commenters that in this limited situation it is unnecessary for the examination chair, which complies with the technical requirements in M302, to also have...
-
Emphasis on Limitations Instead of Outcomes
National Board of Medical Examiners, Gonzales v. National Board of Medical Examiners, and Wong v. Regents of University of California....
-
M302 Diagnostic Equipment used by Patients in a Seated Position (Section-by-Section Analysis)
M302 in the final rule establishes the technical criteria for diagnostic equipment used by patients in a seated position such as examination chairs....
-
M305.3 Standing Supports
Standing supports shall provide continuous support throughout use of the diagnostic equipment and shall comply with M305.3.