A transfer surface shall be provided and shall comply with M302.2.
Search Results "Exam Chair"
-
M302.2 Transfer Surface
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2.
-
M302.2.2 Size
The transfer surface shall be 21 inches (610 mm) wide minimum and 15 inches (381 mm) deep minimum. Advisory M302.2.2 Size. The size requirements in this section apply only to the...
-
M302.2.3 Size
Transfer surfaces shall be 21 inches (610 mm) wide minimum and 17 inches (430 mm) deep minimum. The size of transfer surfaces shall be measured from center points of their opposing...
-
M302.2.3 Transfer Sides
The transfer surface shall be located to provide options to transfer from a mobility device onto one short side (depth) and one long side (width) of the surface. Each transfer side...
-
M302.2.4 Transfer Sides
Options to transfer from a mobility device shall be provided on two adjoining sides of transfer surfaces. EXCEPTION: Options to transfer to or from a mobility device onto...
-
M302.2.5 Unobstructed Transfer
Each transfer side complying with M302.2.4 shall provide unobstructed access to transfer surfaces. EXCEPTIONS: 1. Obstructions no more than 3 inches (75 mm) deep shall be...
-
M302.3 Supports
Transfer supports, leg supports and reclining surfaces shall comply with M302.3.
-
M302.3 Supports
Transfer supports, armrests, and reclining surfaces shall comply with M302.3.
-
M302.3.1 Transfer Supports
Transfer supports shall be provided for use with transfer sides required by M302.2.4 and shall comply with M305.2.1.1, M305.2.2.1, and M305.2.3 through M305.2.8.
-
M302.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M302.2.3 and shall comply with M305.2.
-
M302.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and comply with M305.4.
-
M302.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided and shall comply with M305.5.
-
M302.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided. Where the incline of the back support can be modified while in use, head and...
-
M302.4 Lift Compatibility
Diagnostic equipment shall be usable with portable patient lifts and, when in use with such lifts, shall comply with M302.4.1 or M302.4.2. EXCEPTION: Where fixed overhead patient...
-
M302.4 Lift Compatibility
Diagnostic equipment shall be usable with a patient lift and shall comply with M302.4.1 or M302.4.2. EXCEPTION: Where diagnostic equipment meets the requirements of M303 and...
-
M302.4.1 Clearance in Base
The base of the equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum...
-
M302.4.1 Clearance in Base
The base of the diagnostic equipment shall provide a clearance 39 inches (990 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep...
-
M302.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...
-
M302.4.2 Clearance Around Base
The base of the diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of...
-
Features of an Accessible Examination Room
A clear floor space, 30” X 48” minimum, adjacent to the exam table and adjoining accessible route make it possible to do a side transfer. 2....
-
M305 Supports
-
M305 Supports
-
M305.1 General
Supports shall comply with M305.