and accessibility consultants) recommended retaining the proposed requirement in the MDE NPRM of 48 inches minimum, raising concerns that the size of the rooms in which the diagnostic equipment...
Search Results "Examination Equipment"
-
M303.2.3 Depth (Section-by-Section Analysis)
-
M304.1 General
Diagnostic equipment used by patients in a standing position shall comply with M304....
-
M303.2.1 Orientation (Section-by-Section Analysis)
The MDE NPRM proposed to require wheelchair spaces to be designed so that a patient in a wheelchair using diagnostic equipment would be oriented in the same direction that other non-wheelchair...
-
Transfer Surface (M301.2 and M302.2)
The transfer surface is the part of the diagnostic equipment onto which patients who use mobility devices or aids transfer when moving onto and off of the equipment (see defined terms in...
-
M303.2.1 Orientation
Wheelchair spaces shall be designed so that a patient seated in a wheelchair orients in the same direction that a patient not seated in a wheelchair orients when the diagnostic equipment...
-
M303.2.1 Orientation
Wheelchair spaces shall be designed so that a patient seated in a wheelchair orients in the same direction that a patient not seated in a wheelchair orients when the diagnostic equipment...
-
SELLING OR SERVING SPACE
The rearrangement of temporary or movable structures, such as furniture, equipment, and display racks is not readily achievable to the extent that it results in a significant loss of selling...
-
403.5 Clearances
EXCEPTION: Within employee work areas, clearances on common use circulation paths shall be permitted to be decreased by work area equipment provided that the decrease is essential to the...
-
403.5 Clearances
EXCEPTION: Within employee work areas, clearances on common use circulation paths shall be permitted to be decreased by work area equipment provided that the decrease is essential to the...
-
Section 1193.33 Information, documentation and training [1193.25 in the NPRM] (Section-by-Section Analysis)
This is especially true for good technical support, if persons with disabilities are to receive adequate information on how to use the new accessibility features of telecommunications equipment...
- Living Earth Crafts Essex™ Pedicure Chair
-
11B-206.2.8 Employee work areas
Common use circulation paths located within employee work areas that are an integral component of work area equipment shall not be required to comply with Section 11B-402....
-
11B-206.2.8 Employee work areas
Common use circulation paths located within employee work areas that are an integral component of work area equipment shall not be required to comply with Section 11B-402....
- The Dwarf Athletic Association of America
-
Appendix C to Part 1194—Functional Performance Criteria and Technical Requirements
[See subsections ...]
-
APPENDIX C TO PART 1194 – FUNCTIONAL PERFORMANCE CRITERIA AND TECHNICAL REQUIREMENTS
[See subsections ...]
-
5. Exception from the Requirements of M302 for Weight Scales with Integral Seats
The MDE NPRM proposed that this type of diagnostic equipment would have to meet the technical requirements of M302 (diagnostic equipment used by patients in the seated position) and M303...
-
11B-206.2.8 Employee work areas
Common use circulation paths located within employee work areas that are an integral component of work area equipment shall not be required to comply with Section 11B-402. 3....
-
Section 36.211 Maintenance of Accessible Features (Section-by-Section Analysis)
Section-by-Section Analysis) Section 36.211 of the 1991 title III regulation provides that a public accommodation must maintain in operable working condition those features of facilities and equipment...
-
M303.1 General
Diagnostic equipment used by patients seated in a wheelchair shall comply with M303....
-
M303.1 General
Diagnostic equipment used by patients seated in a wheelchair shall comply with M303....
-
Examination Chair
Diagnostic equipment with a seat in which a patient typically is positioned with buttocks approximately parallel to the ground and shins approximately perpendicular to the ground....
-
M303.2.4.2.2 Height
25 inches (635 mm) maximum at 9 inches (230 mm) above the floor and at a depth of 8 inches (205 mm) minimum at 27 inches (685 mm) above the floor measured from the leading edge of the equipment...
-
B. M301 Diagnostic Equipment Used by Patients in a Supine, Prone, or Side-lying Position and M302 Diagnostic Equipment Used by Patients in a Seated Position
In the final rule M301 and M302 provide the technical requirements for diagnostic equipment used in the supine, prone, or side-lying position, and diagnostic equipment used by patients in...