In the MDE NPRM the Access Board proposed to place the requirements for head and back support for diagnostic equipment used by patients in the seated position in M302....
Search Results "Patient in Seated Position"
-
M302.3.3 Head and Back Support (Section-by-Section Analysis)
-
M302.4 Lift Compatibility (Section-by-Section Analysis)
The MDE NPRM proposed to require that diagnostic equipment used by patients in the seated position be usable with a patient lift and comply with either the proposed clearance in base (proposed...
-
M305 Supports (Section-by-Section Analysis)
M305 in the final rule provides the technical requirements for transfer supports, standing supports, leg supports, and head and back supports. Transfer supports are required for...
-
M305.1 General (Section-by-Section Analysis)
This is an introductory section.
-
M305.2 Transfer Supports (Section-by-Section Analysis)
This is an introductory section. As discussed above in Section IV.E.1. (Significant Changes – Transfer Supports), the Access Board strengthened the transfer support requirements and...
-
M305.2.1 Location (Section-by-Section Analysis)
End transfer supports cover diagnostic equipment used by patients in the supine, prone, or side-lying position with end transfer surfaces, M301.2.3.1 in the final rule, and all diagnostic...
-
M305.2.2 Length (Section-by-Section Analysis)
The Committee explained that the transfer support length provides the gripping surface for the patient to grasp or maintain balance while transferring....
-
M305.2.3 Height (Section-by-Section Analysis)
In the MDE NPRM there was no specific requirement regarding the height of the transfer support, only that it be “within reach” of the patient (proposed M305.2.1)....
-
M305.2.4 Cross Section (Section-by-Section Analysis)
The proposed rule did not provide specific requirements for the cross section of transfer supports. However, in the MDE NPRM preamble, the Access Board noted that it was considering...
-
M305.2.5 Surface Hazards (Section-by-Section Analysis)
were submitted on this issue, but the MDE Advisory Committee voiced concern about surface hazards stating, “gripping surface configurations must provide an effective and safe surface for patients...
-
M305.2.6 Gripping Surfaces (Section-by-Section Analysis)
The MDE Advisory Committee explained that: [t]ransfer supports may contain elements to provide structural support or prevent patient entrapment....
-
M305.2.7 Clearance (Section-by-Section Analysis)
In the MDE NPRM, the Access Board did not provide any specific requirements for clearances around the transfer support. However, in the preamble to the MDE NPRM the Access Board...
-
M305.2.8 Fittings (Section-by-Section Analysis)
Another commenter (manufacturer) pointed out that bed rails, which are common on hospital beds, require a latched position and an unlatched position, which allows them to rotate in their...
-
Figure M3: Overview of M302 Diagnostic Equipment for Use by a Patient in a Seated Position
-
3.1 Getting Weighed
An accessible scale benefits more than just patients who use wheelchairs....
-
M302 Diagnostic Equipment Used by Patients in Seated Position
-
M302 Diagnostic Equipment Used by Patients in Seated Position
-
3.2 Getting on the Table
Additional helpful features include an extra-wide top surface and higher weight capacity for larger people, adjustable side rails, positioning straps, and foot and leg supports that can...
-
M302.1 General
Where diagnostic equipment is used by patients in a seated position, it shall comply with M302....
-
M302.1 General
Diagnostic equipment that supports patients in a seated position shall comply with M302. ...
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2.
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2.
-
M302.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer surface....
-
M302.2.1 Adjustability
A low transfer position at a height of 17 inches (430 mm) minimum and 19 inches (485 mm) maximum; B. A high transfer position at 25 inches (635 mm); and C....