The proposed standards contain minimum technical criteria to ensure that medical diagnostic equipment including examination tables, examination chairs, weight scales, mammography equipment...
Search Results "Strength Training Equipment"
-
1.3.1 Overview of NPRM
-
M301.4 Lift Compatibility
Diagnostic equipment shall be usable with a patient lift and shall comply with M301.4.1 or M301.4.2....
-
Service
Train staff to ask people with disabilities if they need assistance using hospitality language and practices. Keep lowered check-in counter clear....
-
B. Summary of Major Provisions and Organization of Technical Criteria
The Access Board has divided the MDE Standards into separate technical criteria based on how the diagnostic equipment is used by the patient: (1) supine, prone, or side lying position (M301...
-
M303.2.4.1 Toe Clearance
Toe clearance shall be provided at a height of 9 inches (230 mm) above the floor to a depth of 6 inches (150 mm) maximum.
-
M303.2.4.2 Knee Clearance
Knee clearance shall be provided at a depth of 11 inches (280 mm) minimum and 25 inches (635 mm) maximum at 9 inches (230 mm) above the floor and at a depth of 8 inches (205 mm)...
-
Internet Postings (Advisory Guidance)
Internet Postings 15. The fastest growing way to obtain information about a product is through use of the Internet, and specifically the World Wide Web. However, many Internet...
-
Operable Parts
Components of diagnostic equipment that are used by the patient to activate, deactivate, or adjust the equipment....
-
Operable Parts
A component of diagnostic equipment that is used by the patient to activate, deactivate, or adjust the equipment....
- GPK UPPERTONE Unassisted Muscle Strengthening System for Quadriplegics
-
5.4.6 Transfer Support Gripping Surface Cross Section and Clearances for M301, M302, and M305.2
The Access Board is considering whether the above cross section dimensions would be appropriate for the gripping surfaces of transfer supports on diagnostic equipment used by patients in...
-
M305 Supports
The technical criteria for transfer supports are discussed under M301 Diagnostic Equipment Used by Patients in Supine, Prone, or Side-Lying Position and M302 Diagnostic Equipment Used by...
-
II. Rulemaking History
Section 510 of the Rehabilitation Act requires the Access Board to issue standards for medical diagnostic equipment to ensure such equipment is accessible to, and usable by, individuals...
-
M301.3.1 Transfer Supports (Section-by-Section Analysis)
The MDE NPRM proposed to require transfer supports to be provided for use with transfer sides on diagnostic equipment used by patients in the supine, prone, or side-lying position, and that...
-
1004.1 Clear Floor Space
Clear floor or ground spaces required at exercise machines and equipment shall be permitted to overlap. Advisory 1004.1 Clear Floor Space....
-
1004.1 Clear Floor Space
Clear floor or ground spaces required at exercise machines and equipment shall be permitted to overlap. Advisory 1004.1 Clear Floor Space....
-
II. Systemic Considerations
Training criminal justice personnel, conducting reviews of policies and procedures, and collaborating with mental health and disability services providers are three ways to achieve that...
-
11B-1004.1 Clear floor space
Clear floor or ground spaces required at exercise machines and equipment shall be permitted to overlap. [2010 ADAS] Advisory 1004.1 Clear Floor Space. ...
-
1004.1 Clear Floor Space
Clear floor or ground spaces required at exercise machines and equipment shall be permitted to overlap. Advisory 1004.1 Clear Floor Space....
-
11B-1004.1 Clear floor space
Clear floor or ground spaces required at exercise machines and equipment shall be permitted to overlap. [2010 ADAS] Advisory 1004.1 Clear Floor Space....
-
11B-1004.1 Clear floor space
Clear floor or ground spaces required at exercise machines and equipment shall be permitted to overlap. Advisory 11B-1004.1 Clear floor space....
- Detecto 485 Series Stationary Wheelchair Scale
-
M301.2.3.1 End Transfer Surface
EXCEPTION: Transfer surfaces for imaging equipment with bores shall be permitted to be 21 inches (535 mm) wide minimum but shall not be permitted to be less than the full width of the examination...
-
M301.2.3.2 Side Transfer Surface
EXCEPTION: Transfer surfaces for imaging equipment with bores shall be permitted to be 21 inches (535 mm) wide minimum but shall not be permitted to be less than the full width of the examination...