(at Diagnostic Equipment for Use by a Patient in a Supine, Prone, or Side-Lying Position)
Search Results "Procedure Table"
-
Figure M2: M301.4 Lift Compatibility
-
M301.4.1 Clearance in Base
The base of the equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum...
-
5.3.2 Transfer Sides Recommendations for M301
., foot of an examination table)....
-
1.1 Recognizing the Barriers
training about disability or accessibility issues The staff and some members have negative attitudes about people with disabilities Some of the facility’s policies and procedures...
-
Your ADA Obligations
facilities; Remove architectural or structural communication barriers in existing facilities where it is readily achievable to do so; Make reasonable modifications in policies and procedures...
- Feeling Safe, Being Safe Webcast Training: Community Supports
- Widgit First Response Communication Book
-
Release Planning
Release Planning: The settlement agreement with Hinds County, Mississippi requires the jail to notify community mental health providers when releasing a prisoner with serious mental...
-
Disability Specific Supplies for Emergency Kits
Edition 2.0, 2016 By June Isaacson Kailes, Disability Policy Consultant and Associate Director Harris Family Center for Disability and Health Policy at Western University of...
-
III-4.2000 Reasonable modifications
[See subsections ...]
-
M305.4 Leg Supports (Section-by-Section Analysis)
The MDE Advisory Committee agreed that, for procedures that use stirrups and require the leg to be stable, there must be a method to support the patient’s legs....
-
M305.5 Head and Back Support (Section-by-Section Analysis)
One manufacturer asserted that this requirement was ambiguous and that he had to read it multiple times to understand it; however, this commenter also indicated that the tables it currently...
-
Clear Floor and Turning Space Inside Examination Rooms
The exam table must have sufficient clear floor space next to it so that an individual using a wheelchair can approach the side of the table for transfer onto it....
-
a. Transfer Surface Adjustability
The MDE NPRM proposed that the same transfer surface height range of 17 inches minimum to 19 inches maximum be applied to both diagnostic equipment used in the supine, prone, or...
-
11B-249.1.2
Newly constructed commercial places of public amusement shall provide no fewer than one adult changing facility in compliance with Section 11B-813. ETA Editor's Note The...
-
604.8.1.1 Size
This space cannot be obstructed by baby changing tables or other fixtures or conveniences, except as specified at 604.3.2 (Overlap)....
-
604.8.1.1 Size
This space cannot be obstructed by baby changing tables or other fixtures or conveniences, except as specified at 604.3.2 (Overlap)....
-
604.8.1.1 Size
This space cannot be obstructed by baby changing tables or other fixtures or conveniences, except as specified at 604.3.2 (Overlap)....
-
How to Measure Transfer Surface Height
Many medical examination rooms are equipped with fixed height examination tables, with a typical 32-inch seat height....
- Living Earth Crafts Cloud 9 Spa Treatment Table
-
M301.4.2 Clearance Around Base
The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...
-
M301.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...
-
M301.4.2 Clearance Around Base
[See subsections ...]
-
§382.125 What procedures do carriers follow when wheelchairs, other mobility aids, and other assistive devices must be stowed in the cargo compartment?
[See subsections ...]