5.4.10 Transfer Support Fittings for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. The transfer supports shall not rotate when locked in place for patient...
Search Results "Built-In Table"
Commonly Searched Documents
-
5.4.10 Transfer Support Fittings for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
-
18 Inch Low Height
18 Inch Low Height As our discussions progressed on the low height, whether it should be 17 inches, 18 inches or 19 inches; a compromise position of 18 inches was suggested. Some...
-
Height – M301.2.1 & M302.2.1
Height – M301.2.1 & M302.2.1 The subcommittee discussed the increments for adjustability and decided that infinite adjustability is the appropriate standard as both electronic...
-
M305.2.3 Fittings
Transfer supports shall not rotate within their fittings.
- Hussey Seating Company Roll-A-Way Seating
- Irwin Seating Company ADA Movable Base
- Hussey Seating Company Skid-Mounted and Companion Chairs
-
Accessible Medical Equipment
Importance of Accessible Examination Tables, Chairs and Weight Scales http://www.hfcdihp.org/products Exam Room Selection for Accessible Examination Tables & Chairs PDF , Word...
-
5.1 Transfer Surface Height Recommendations
This height is also critical for patients who use mobility aids such as walkers and canes and may find it difficult to get up onto or down from an examination chair or table or imaging bed...
-
Kompan, Inc. Sandboxes and Water play - ELE400101 - Ministudio 1
Requires the Premium User plan or above. (Login or upgrade your account!)
-
Kompan, Inc. M7000 - Red House Playhouse
Requires the Premium User plan or above. (Login or upgrade your account!)
-
Kompan, Inc. GSP000100 - Basic Play House
Requires the Premium User plan or above. (Login or upgrade your account!)
-
I. BACKGROUND
Litigation include, among other things, not being placed in an accessible inpatient room; being examined on inaccessible equipment which required them to be lifted onto an examination table...
-
36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment NPRM
The proposed standards contain minimum technical criteria to ensure that medical diagnostic equipment, including examination tables, examination chairs, weight scales, mammography equipment...
-
M305.2.1 Location
Transfer supports shall comply with M305.2.1.1 or M305.2.1.2 and shall be located 1 ½ inches (38 mm) maximum measured horizontally from the plane defined by the nearest edge of the...
-
M305.2.4 Cross Section
Transfer supports shall have a cross section complying with 305.2.4.1 or 305.2.4.2.
-
M305.2.3 Height
During use, the tops of transfer support gripping surfaces shall be 6 inches (150 mm) minimum and 19 inches (485 mm) maximum higher than the top of the associated uncompressed...
-
M305.2.6 Gripping Surface
Transfer support gripping surfaces shall be continuous along their length and shall not be obstructed along their tops or sides. The bottoms of transfer support gripping surfaces...
-
M305.2.6 Gripping Surfaces (Section-by-Section Analysis)
The proposed rule did not provide any specific requirements regarding gripping surfaces on transfer supports. However, in the MDE NPRM preamble the Access Board repeatedly noted that...
-
M305.2.4 Cross Section (Section-by-Section Analysis)
The proposed rule did not provide specific requirements for the cross section of transfer supports. However, in the MDE NPRM preamble, the Access Board noted that it was considering...
-
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302 Description: Transfer supports may contain elements to provide structural support or prevent...
-
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. Transfer supports should be required on both sides of the transfer surface and...
-
5.4.5 Transfer Support Position Recommendation for M301 and M302
5.4.5 Transfer Support Position Recommendation for M301 and M302 Description: Transfer support position refers to the placement of the gripping surface within the minimum and...
-
5.4.2 Transfer Support Length for M301 and M302
5.4.2 Transfer Support Length for M301 and M302 Description: Transfer support length applies to the gripping surface of the support, that portion of the support used to grasp or...