The MDE NPRM proposed to restrict the force required to activate operable parts to 5 pounds. The Access Board sought public input on this provision in question 42 on whether the...
Search Results "Strength Training Equipment"
-
M307.4 Operating Force (Section-by-Section Analysis)
-
M307.4 Operating Force
The force required to activate operable parts shall be 5 pounds (22.2 N) maximum.
-
3. Background
[See subsections ...]
-
C201.5 Design, Development and Fabrication (Section-by-Section Analysis)
C201.5 Design, Development and Fabrication (Section-by-Section Analysis) This section proposes a general requirement that telecommunications equipment manufacturers evaluate the accessibility...
-
M101.3 Equivalent Facilitation
., use of a patient lift) are not permitted because they do not provide for independent access to and use of diagnostic equipment by patients with disabilities to the maximum extent possible...
-
5.4.6 Transfer Support Gripping Surface Cross Section for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.6 Transfer Support Gripping Surface Cross Section for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
M305.3 Standing Supports (Section-by-Section Analysis)
M305.3 provides the technical requirements for standing supports which are required on diagnostic equipment covered by M304....
-
Compliance requirements in proposed standards
The proposed standards contain technical criteria for accessible medical diagnostic equipment....
-
5.4.7 Transfer Support Gripping Surface Clearance for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.7 Transfer Support Gripping Surface Clearance for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position Description: Many people who ambulate with a disability need support to maintain...
-
1630.4(b)
(b) Hiring, upgrading, promotion, award of tenure, demotion, transfer, layoff, termination, right of return from layoff, and rehiring;
-
5.8 Wheelchair Spaces for Diagnostic Equipment Used by Patients Seated in a Wheelchair
5.8 Wheelchair Spaces for Diagnostic Equipment Used by Patients Seated in a Wheelchair....
-
Cassette Recordings (Advisory Guidance)
Cassette Recordings 9. Some persons who are blind or who have learning disabilities may require documentation on audio cassettes. Audio materials can be produced commercially or...
-
Large Print (Advisory Guidance)
Large Print 7. Persons with low vision may require documentation to be provided in large print. Large print documents can easily be produced using a scalable font from any good...
-
Electronic Text (Advisory Guidance)
Electronic Text 5. People who are blind or have low vision and who have access to computers may be able to use documents in electronic form. Electronic text must be provided in...
-
5.2.1.2 Transfer Surface Depth
The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception of imaging equipment....
-
C201.5 Design, Development, and Fabrication
Telecommunications equipment manufacturers shall evaluate the accessibility, usability, and interoperability of ICT during its product design, development, and fabrication....
-
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.3 Transfer Support Height for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.3 Transfer Support Height for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
M302.3.2 Leg Supports (Section-by-Section Analysis)
The MDE NPRM did not propose to require stirrups to provide a method of supporting, positioning, and securing the patient’s legs for diagnostic equipment used by patients in the seated position...
-
M301.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination surface...
-
A. Chapter 1: Application and Administration (Section-by-Section Analysis)
The Access Board made a few editorial changes to some of the provisions, and added one provision M101.3 Existing Diagnostic Equipment, which is discussed below....
-
M301.4.2 Clearance Around Base
The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
5.2.1.1 Transfer Surface Width Recommendations for M301
Currently, the only equipment existing with a 30-inch transfer surface width is bariatric equipment. At the Committee’s December 2012 meeting, Dr....