[See subsections ...]
Search Results "Telecommunications Equipment Manufacturer"
-
3. Background
-
M101.3 Equivalent Facilitation
., use of a patient lift) are not permitted because they do not provide for independent access to and use of diagnostic equipment by patients with disabilities to the maximum extent possible...
-
5.4.6 Transfer Support Gripping Surface Cross Section for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.6 Transfer Support Gripping Surface Cross Section for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
If a more recent edition of ANSI/BHMA Standards for automated doors is issued should that be followed instead of those referenced?
The specific editions of the ANSI/BHMA Standards referenced by the Standards are to be followed (although the Access Board intends to update references to keep them current)....
-
M305.2.3 Height (Section-by-Section Analysis)
Three commenters (a manufacturer, a medical association, and a disability rights organization) did not support the proposal....
-
5.4.7 Transfer Support Gripping Surface Clearance for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.7 Transfer Support Gripping Surface Clearance for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
III. Summary of Comments
The manufacturers provided a great deal of information on what types of accessible equipment is currently on the market, what the providers are requesting for accessible equipment, and the...
-
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position Description: Many people who ambulate with a disability need support to maintain...
-
M307 Operable Parts
Board is considering requiring in the final standards for operable parts used by patients that require fine grips: a) What would be the incremental costs for the design or redesign and manufacture...
-
5.8 Wheelchair Spaces for Diagnostic Equipment Used by Patients Seated in a Wheelchair
5.8 Wheelchair Spaces for Diagnostic Equipment Used by Patients Seated in a Wheelchair....
- KRANKcycle® Informational Video with Mike Michaels
-
5.2.1.2 Transfer Surface Depth
The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception of imaging equipment....
-
A. Access Board
Rehabilitation Act (29 U.S.C. 792) for buildings and facilities; Section 508 of the Rehabilitation Act (29 U.S.C. 794d) for electronic and information technology; and Section 255 of the Telecommunications...
-
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.3 Transfer Support Height for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.3 Transfer Support Height for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
M301.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination surface...
-
A. Chapter 1: Application and Administration (Section-by-Section Analysis)
The Access Board made a few editorial changes to some of the provisions, and added one provision M101.3 Existing Diagnostic Equipment, which is discussed below....
-
M301.4.2 Clearance Around Base
The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
5.2.1 Transfer Surface Size Recommendations for M301
5.2.1 Transfer Surface Size Recommendations for M301 Description: The transfer surface is the part of the diagnostic equipment onto which patients who use mobility devices or aids transfer...
-
M101.2 Application
The standards shall be applied to diagnostic equipment based on the patient positions that the equipment is designed to support....
-
M305.2.2 Length (Section-by-Section Analysis)
extend horizontally along the transfer surface to within three inches, to allow for manufacturing tolerances....
- Taking your first Spin on the Johnny G KRANKcycle® by Matrix
- How to File a Complaint with the Federal Communications Commission (FCC)
-
Equipment and furniture.
obligations to provide accessible free-standing equipment, the Department was considering adding specific language on equipment. ...
