The length of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve....
Search Results "Standing Support"
-
M305.3.1 Length
-
M305.3.1 Horizontal Position
Where the support is horizontal, the top of the gripping surface shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface....
-
M305.3.1.1 Horizontal Position
The length of gripping surfaces on horizontal standing supports shall be 4 inches (100 mm) minimum except for diagnostic equipment containing a wheelchair space which shall comply with M305.3.1.2...
-
M305.3.1.2 Diagnostic Equipment Containing a Wheelchair Space
On diagnostic equipment containing wheelchair spaces with one entry that also serves as the exit, the length of the gripping surface of horizontal standing supports shall be equal to or...
-
M305.3.1.3 Vertical Position
The length of the gripping surface on vertical standing supports shall be 18 inches (455 mm) minimum....
-
M305.3.2 Vertical Position
Where the support is vertical, it shall be 18 inches (455 mm) minimum in length and the bottom end of the support shall be 34 inches (865 mm) high minimum and 37 inches (940 mm) high maximum...
-
M305.3.2 Height
The height of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve....
-
M305.3.2.1 Horizontal Position
The height of the top of the gripping surface on horizontal standing supports shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface....
-
M305.3.2.2 Vertical Position
The height of the lowest end of the gripping surface on vertical standing supports shall be 34 inches (865 mm) minimum and 37 inches (940 mm) maximum above the standing surface....
-
M305.3.3 Fittings
Standing supports shall not rotate within their fittings....
-
M305.3.3 Fittings
Standing supports shall not rotate within their fittings....
-
Figure M4: Overview of M303 Diagnostic Equipment for Use by a Patient While Seated in a Wheelchair and M304 Diagnostic Equipment for Use by a Patient in a Standing Position (Weight Scale Example)
Where the primary use of the equipment supports more than one patient position, it must comply with the requirements for each use. ...
-
Figure M5: Overview of M303 Diagnostic Equipment for Use by a Patient While Seated in a Wheelchair and 304 Diagnostic Equipment for Use by a Patient in a Standing Position (Mammography Machine Example)
Where the primary use of the equipment supports more than one patient position, it must comply with the requirements for each use. ...
-
Standing Supports (M304.3 and M305.3)
M304.3 would require standing supports to be provided on each side of the standing surface on diagnostic equipment used by patients in a standing position....
-
M305 Supports
M305 provides the technical criteria for transfer supports and standing supports....
-
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position Description: Many people who ambulate with a disability need support to maintain...
-
5.9.1 Standing Supports Recommendation In Reference To Mammography Equipment
5.9.1 Standing Supports Recommendation In Reference To Mammography Equipment The Committee recommends removing the requirement for standing supports on mammography equipment....
-
5.9.2 Standing Supports on Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2 Standing Supports on Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations...
-
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on both sides of platforms,...
-
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on one side of the platform,...
-
Medical Diagnostic Equipment Accessibility Standards Sub-Committee Recommendations Weight Scales
Specifically, section M303, patient positions equipment designed to support seated in a wheelchair....
-
Standing Supports (M304.3 and M305.3)
Standing Supports (M304.3 and M305.3) M304.3 Proposed Provision: Would require standing supports to be provided on each side of the standing surface on diagnostic equipment used by patients...