The base of the diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of...
Search Results "Lift Compatibility"
-
M302.4.2 Clearance Around Base
-
Lift Compatibility (M301.4 and M302.4)
Portable floor lifts have legs with wheels that need to fit under or around the base of the diagnostic equipment....
-
Clearance in Base (M301.4.1 and M302.4.1)
Clearance in the base of the equipment allows the legs of a portable floor lift to fit under the base of the equipment....
-
Clearance Around Base (M301.4.2 and M302.4.2)
Clearance around the base of the equipment allows the legs of a portable floor lift to straddle the base....
-
5.7 Lift Compatibility Recommendations
5.7 Lift Compatibility Recommendations...
-
5.7.1 Lift Compatibility Clearance in Base for M301
5.7.1 Lift Compatibility Clearance in Base for M301 Description: For portable floor lifts to be effective, they must deploy at diagnostic equipment such that the boom of the lifts can...
-
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers (M301)
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers (M301) The Committee recommends the equipment base provide a clearance of 39 inches wide minimum....
-
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment (M301)
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment (M301) The Committee recommends the use of alternative means of access by an overhead lift in place of clearances in...
-
M301.4 Lift Compatibility
M301.4 Lift Compatibility Subcommittee Recommendation: (Also agreed to by the full committee.)...
-
M301.4 Lift Compatibility
M301.4 Lift Compatibility M01.4.1 Clearance in Base Proposed requirement: The base of the equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm)...