5.2.1.4 Transfer Surface Size for Imaging Equipment. The transfer surface should be 28 inches wide minimum and 21 inches deep minimum. The transfer surface should be located such...
Search Results "Horizontal Position"
-
5.2.1.4 Transfer Surface Size for Imaging Equipment
-
M301.2.2 Transfer Surface Size
M301.2.2 Transfer Surface Size Subcommittee Recommendation: (Also agreed to by the full committee). A. The minimum size of the transfer surface is 28 inches wide minimum and 21...
-
5.2.1.2 Transfer Surface Depth
The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception of imaging equipment....
-
11B-411.4.7 Designations and indicators of car control buttons
Designations and indicators of car control buttons shall comply with Section 11B-411.4.7.
-
11B-411.4.7 Designations and indicators of car control buttons
Designations and indicators of car control buttons shall comply with Section 11B-411.4.7.
-
(2) Transfer Surface Location for Diagnostic Equipment Used in the Seated Position
Commenters also raised concerns with the provisions in the MDE NPRM related to transferring to medical diagnostic equipment used by patients in the seated position....
-
M301.3.2 Stirrups
Where stirrups are provided, they shall provide a method of supporting, positioning, and securing the patient’s legs....
-
Gripping Surface Cross Section and Clearances M301, M302, and M305.2
Gripping Surface Cross Section and Clearances M301, M302, and M305.2 Gripping surface for transfer supports shall be free of sharp or abrasive elements and shall have rounded...
-
M305.2.2.1 End Transfer Supports
Transfer supports shall be positioned along 13 ½ inches (345 mm) minimum of the depth of the transfer surface....
-
M302.1 General
Where diagnostic equipment is used by patients in a seated position, it shall comply with M302....
-
M301.2.1 Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface height range for diagnostic equipment used by patients in the supine, prone, or side-lying position of 17 inches minimum and 19 inches maximum....
-
5.4.3 Transfer Support Height for M301 and M302
The purpose of a minimum height is to ensure that the gripping surface is at an adequate height to facilitate positioning during a transfer....
-
M301.3.2 Leg Supports (Section-by-Section Analysis)
In the MDE NPRM, the Access Board proposed to place the requirements for stirrups on diagnostic equipment used by patients in the supine, prone, or side-lying position in M301....
-
M301.3.3 Head and Back Support (Section-by-Section Analysis)
In the MDE NPRM the Access Board proposed to place the requirements for head and back support for diagnostic equipment used by patients in the supine, prone, or side-lying position in M301...
-
M305.5 Head and Back Support (Section-by-Section Analysis)
The MDE NPRM proposed to require diagnostic equipment used by patients in the supine, prone, or side-lying position and the seated position that can be adjusted to a reclined position to...
-
M304.2.2 Standing Supports
other shall be permitted to provide one standing support on one side of the standing surface provided that the standing support complies with the requirements for standing supports in a horizontal...
-
1007.4 Elevators
The elevator shall be accessed from either an area of refuge complying with Section 1007.6 or a horizontal exit....
-
M305.2.2.2 Side Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.2 shall be 28 inches (710 mm) long minimum and shall be positioned along the width of transfer surfaces....
-
M301.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided and shall comply with M305.5....
-
1003.2.13.2 Enclosed exit stairways
handrails and shall either incorporate an area of refuge within an enlarged floor-level landing or shall be accessed from either an area of refuge complying with Section 1003.2.13.5 or a horizontal...
-
5.4.6 Transfer Support Gripping Surface Cross Section and Clearances for M301, M302, and M305.2
whether the above cross section dimensions would be appropriate for the gripping surfaces of transfer supports on diagnostic equipment used by patients in a supine, prone, or side-lying position...
-
M301.2.3 Size (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface size for diagnostic equipment used in the supine, prone, or side-lying position of 30 inches wide and 15 inches deep minimum....
-
M101.2 Application
Sections M301 through M304 shall be applied to diagnostic equipment, based on the patient positions that the equipment supports, during patient transfer and diagnostic use....
-
5.1.2 Transfer Surface High Height Recommendations for M301 and M302
5.1.2 Transfer Surface High Height Recommendations for M301 and M302 The Committee recommends an adjustable height range with an upper height measuring at least 25...