Search Results "Lift Compatibility"
-
5.7 Lift Compatibility
-
5.7.1 Lift Compatibility Clearance in Base for Equipment Used by Patient in Supine, Prone, or Side-Lying Position
5.7.1 Lift Compatibility Clearance in Base for Equipment Used by Patient in Supine, Prone, or Side-Lying Position....
-
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers. The base of the equipment must provide a clearance of 39 inches wide minimum....
-
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment....
-
B. M301 Diagnostic Equipment Used by Patients in a Supine, Prone, or Side-lying Position and M302 Diagnostic Equipment Used by Patients in a Seated Position
Therefore, the Significant Changes Section addresses the transfer surface and lift compatibility requirements for M301 and M302 together....
-
3. Lift Compatibility Exception
The proposed rule specified base clearance requirements to ensure lift compatibility (M301.4 and M302.4, respectively)....
-
M301.4 Lift Compatibility (Section-by-Section Analysis)
Additionally, question 27 in the MDE NPRM preamble requested input on whether the final rule should provide an exception from the lift compatibility requirements where the diagnostic equipment...
-
M301.4.1 Clearance in Base (Section-by-Section Analysis)
This commenter explained a preference for fixed overhead lifts....
-
M301.4.2 Clearance Around Base (Section-by-Section Analysis)
One commenter recommended clarifying that the exam table must be compatible with a patient lift and meet the six-inch clearance, but not when the table is at its lowest level for independent...
-
M302.4 Lift Compatibility (Section-by-Section Analysis)
This requirement is identical to the lift compatibility requirement for diagnostic equipment used by patients in the supine, prone, or side-lying position, and is discussed in the Section-by...
-
Figure M2: M301.4 Lift Compatibility
(at Diagnostic Equipment for Use by a Patient in a Supine, Prone, or Side-Lying Position)
-
M301.4.1 Clearance in Base
-
M301.4.2 Clearance Around Base
-
M301.4 Lift Compatibility
Diagnostic equipment shall be usable with a patient lift and shall comply with M301.4.1 or M301.4.2....
-
M301.4 Lift Compatibility
Diagnostic equipment shall be usable with portable patient lifts and, when in use with such lifts, shall comply with M301.4.1 or M301.4.2....
-
M301.4.1 Clearance in Base
The base of the equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum...
-
M301.4.1 Clearance in Base
The base of diagnostic equipment shall provide a clearance 39 inches (990 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep...
-
M301.4.2 Clearance Around Base
The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...
-
M301.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...
-
M302.4 Lift Compatibility
Diagnostic equipment shall be usable with portable patient lifts and, when in use with such lifts, shall comply with M302.4.1 or M302.4.2. ...
-
M302.4 Lift Compatibility
Diagnostic equipment shall be usable with a patient lift and shall comply with M302.4.1 or M302.4.2....
-
M302.4.1 Clearance in Base
The base of the equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum...
-
M302.4.1 Clearance in Base
The base of the diagnostic equipment shall provide a clearance 39 inches (990 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep...
-
M302.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...