M305 provides the technical requirements for supports on medical diagnostic equipment....
Search Results "Horizontal Standing Support"
-
E. M305 Supports
-
M305.3.1 Length (Section-by-Section Analysis)
In the MDE NPRM, the Access Board proposed a gripping surface length of four inches minimum for horizontal standing supports. No public comments were submitted on this requirement....
-
M305.3.1 Length
The length of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve....
-
M305.3.1 Horizontal Position
Where the support is horizontal, the top of the gripping surface shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface....
-
M305.3.1.1 Horizontal Position
The length of gripping surfaces on horizontal standing supports shall be 4 inches (100 mm) minimum except for diagnostic equipment containing a wheelchair space which shall comply with M305.3.1.2...
-
M305.3.1.2 Diagnostic Equipment Containing a Wheelchair Space
On diagnostic equipment containing wheelchair spaces with one entry that also serves as the exit, the length of the gripping surface of horizontal standing supports shall be equal to or...
-
M305.3.2.1 Horizontal Position
The height of the top of the gripping surface on horizontal standing supports shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface....
-
Figure M4: Overview of M303 Diagnostic Equipment for Use by a Patient While Seated in a Wheelchair and M304 Diagnostic Equipment for Use by a Patient in a Standing Position (Weight Scale Example)
Where the primary use of the equipment supports more than one patient position, it must comply with the requirements for each use. ...