Figure M1: Overview of M301 Diagnostic Equipment for Use by a Patient in a Supine, Prone, or Side-Lying Position
-
5.7.1 Lift Compatibility Clearance in Base for Equipment Used by Patient in Supine, Prone, or Side-Lying Position
5.7.1 Lift Compatibility Clearance in Base for Equipment Used by Patient in Supine, Prone, or Side-Lying Position....
-
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers. The base of the equipment must provide a clearance of 39 inches wide minimum....
-
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment....
-
3. Lift Compatibility Exception
The proposed rule specified base clearance requirements to ensure lift compatibility (M301.4 and M302.4, respectively)....
-
M301.4 Lift Compatibility (Section-by-Section Analysis)
in base (proposed M301.4.1) or clearance around base (proposed M301.4.2) technical requirements....
-
M301.4.1 Clearance in Base (Section-by-Section Analysis)
The MDE NPRM proposed a clearance in the base of 44 inches wide minimum, 6 inches high minimum measured from the floor, and 36 inches deep minimum measured from the edge of the examination...
-
M301.4.1 Clearance in Base
-
Figure M3: Overview of M302 Diagnostic Equipment for Use by a Patient in a Seated Position
-
M301.4.1 Clearance in Base
The base of the equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum measured...
-
M301.4.1 Clearance in Base
The base of diagnostic equipment shall provide a clearance 39 inches (990 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum measured...
-
M302.4.1 Clearance in Base
The base of the equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum measured...
-
M302.4.1 Clearance in Base
The base of the diagnostic equipment shall provide a clearance 39 inches (990 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum...
-
Clearance in Base (M301.4.1 and M302.4.1)
Clearance in the base of the equipment allows the legs of a portable floor lift to fit under the base of the equipment....
-
5.7.1 Lift Compatibility Clearance in Base for M301
5.7.1 Lift Compatibility Clearance in Base for M301 Description: For portable floor lifts to be effective, they must deploy at diagnostic equipment such that the boom of the lifts can...
-
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers (M301)
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers (M301) The Committee recommends the equipment base provide a clearance of 39 inches wide minimum....
-
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment (M301)
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment (M301) The Committee recommends the use of alternative means of access by an overhead lift in place of clearances in...