Figure M1: Overview of M301 Diagnostic Equipment for Use by a Patient in a Supine, Prone, or Side-Lying Position
-
3. Lift Compatibility Exception
The proposed rule specified base clearance requirements to ensure lift compatibility (M301.4 and M302.4, respectively)....
-
M301.4 Lift Compatibility (Section-by-Section Analysis)
in base (proposed M301.4.1) or clearance around base (proposed M301.4.2) technical requirements....
-
M301.4.2 Clearance Around Base (Section-by-Section Analysis)
The MDE NPRM proposed certain requirements to provide clearance around the base of the diagnostic equipment to allow the legs of the portable floor lift to straddle the base of the diagnostic...
-
M301.4.2 Clearance Around Base
-
Figure M3: Overview of M302 Diagnostic Equipment for Use by a Patient in a Seated Position
-
M301.4.2 Clearance Around Base
The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
M301.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination surface...
-
M302.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination surface...
-
M302.4.2 Clearance Around Base
The base of the diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
Clearance Around Base (M301.4.2 and M302.4.2)
Clearance around the base of the equipment allows the legs of a portable floor lift to straddle the base....