5.1 Transfer Surface Height
Search Results "Adjustability"
-
5.1 Transfer Surface Height
-
5.1.1 Transfer Surface Height Adjustability
5.1.1 Transfer Surface Height Adjustability. The transfer surface should have an adjustable height in small virtually continuous increments....
-
5.1.2 Transfer Surface High Height
The transfer surface should have an adjustable height range with an upper height measuring at least 25 inches....
-
5.1.3 Transfer Surface Low Height
5.1.3 Transfer Surface Low Height. No agreement on final low height
-
5.1.4 Transfer Surface Height Measurement
5.1.4 Transfer Surface Height Measurement. The transfer surface height should be measured from the floor to the highest point on the seat in an uncompressed state, inclusive of...
-
a. Transfer Surface Adjustability
The Board considered it likely that diagnostic equipment would be adjustable in height to serve practitioners’ needs however, the transfer surface could be fixed within the proposed height...
-
(1) Adjustability: Minimum High Transfer Height
In the preamble to the MDE NPRM, the Access Board sought comment in question 14 on whether the final rule should require an adjustable height range of 17 inches to 25 inches; whether equipment...
-
(2) Adjustability: Minimum Low Transfer Height
The Access Board received many comments from disability rights organizations, individuals, accessibility consultants, and a health care provider supporting the need for lower height adjustable...
-
(3) Adjustability: Transfer Surface Intermediate Heights
The MDE Advisory Committee recommended adjustable height in small, virtually continuous increments....
-
(4) Adjustability: Method of Measurement
The MDE NPRM proposed that the measurement of the height of the transfer surface for both diagnostic equipment used in the supine, prone, or side-lying position and diagnostic...
-
4. Breast Platform Adjustability
Therefore, the final rule requires at M303.4.1 that breast platforms have a minimum low height of 26 inches, a minimum high height of 42 inches, and be continually adjustable between the...
-
M301.2.1 Adjustability (Section-by-Section Analysis)
(Significant Changes – Transfer Surface Adjustability), in the final rule the Access Board has renamed this provision and now requires the transfer surface height to be adjustable to: (1...
-
M301.2.2 Sunset Provision (Section-by-Section Analysis)
(Significant Changes – Transfer Surface Adjustability), this is a new provision that was added to the final rule in conjunction with the new requirement of a low height range in M301.2.1...
-
M302.2.1 Adjustability (Section-by-Section Analysis)
(Significant Changes – Transfer Surface Adjustability) in the final rule the Access Board has renamed this provision and now requires the transfer surface to be adjustable to: (1) a low...
-
M302.2.2 Sunset Provision (Section-by-Section Analysis)
(Significant Changes – Transfer Surface Adjustability), this is a new provision added to the final rule in conjunction with the new requirement of a low height range in M302.2.1....
-
M303.4.1 Breast Platform Adjustability (Section-by-Section Analysis)
(Significant Changes – Breast Platform Adjustability), the Access Board has revised this provision to require the breast platform to be continually adjustable from a low height of 26 inches...
-
M301.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M301.2.2 Sunset
The low transfer position height, Item A of M301.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register].
-
M302.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer...
-
M302.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M302.2.2 Sunset
The low transfer position height, Item A of M302.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register]. [sic]
-
M303.4.1 Breast Platforms
The height of the breast platform shall be 30 inches (760 mm) high minimum and 42 inches (1065 mm) high maximum above the floor when in use by a patient seated in a wheelchair.
-
M303.4.1 Breast Platform Adjustability
Breast platforms shall be continuously adjustable from a low height of 26 inches (660 mm) to a high height of 42 inches (1065 mm) above the floor....
-
5.1 Transfer Surface Height Recommendations
NPRM Preamble Discussion: The technical criteria allow the height of transfer surfaces to be either fixed or adjustable within the 17 inches minimum and 19 inches maximum range....