Independent Wheelchair Transfers in the Built Environment: How Transfer Setup Impacts Performance Phase 2: Final Report
-
5.1 Transfer Surface Height
5.1 Transfer Surface Height...
-
5.1.1 Transfer Surface Height Adjustability
5.1.1 Transfer Surface Height Adjustability. The transfer surface should have an adjustable height in small virtually continuous increments....
-
5.1.2 Transfer Surface High Height
5.1.2 Transfer Surface High Height. The transfer surface should have an adjustable height range with an upper height measuring at least 25 inches....
-
5.1.3 Transfer Surface Low Height
5.1.3 Transfer Surface Low Height. No agreement on final low height...
-
5.1.4 Transfer Surface Height Measurement
5.1.4 Transfer Surface Height Measurement....
-
M301.2.1 Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface height range for diagnostic equipment used by patients in the supine, prone, or side-lying position of 17 inches minimum and 19 inches maximum....
-
M301.2.2 Sunset Provision (Section-by-Section Analysis)
(Significant Changes – Transfer Surface Adjustability), this is a new provision that was added to the final rule in conjunction with the new requirement of a low height range in M301.2.1...
-
M302.2.1 Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface height range for diagnostic equipment used by patients in a seated position of 17 inches minimum and 19 inches maximum....
-
M302.2.2 Sunset Provision (Section-by-Section Analysis)
(Significant Changes – Transfer Surface Adjustability), this is a new provision added to the final rule in conjunction with the new requirement of a low height range in M302.2.1....
-
M301.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M301.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer surface....
-
M301.2.2 Sunset
The low transfer position height, Item A of M301.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register]....
-
M302.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer surface....
-
M302.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M302.2.2 Sunset
The low transfer position height, Item A of M302.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register]. [sic]...
-
Implication for Standards and Design
The station used in this second phase permitted greater freedom of wheelchair positioning around the transfer surface compared to the first phase station. ...
-
Clear Space Evaluation
As a result subjects positioned themselves in a wide variety of locations about the transfer station. ...
-
Grab Bar and Backrest Evaluation
The grab bars were frequently used as handhelds during the transfer process and for one person it made the difference of being able to do a non-level transfer to the platform....
-
5.1 Transfer Surface Height Recommendations
5.1 Transfer Surface Height Recommendations Description: The transfer height is set above the finished floor surface to permit transfers form mobility devices onto the transfer surface...
-
5.1.1 Transfer Surface Height Adjustability Recommendation for M301 and M302
Adjustable height MDE, such as exam tables, imaging tables and chairs, will make it possible to position the transfer surface near the height of the seat of the mobility device....
-
5.1.2 Transfer Surface High Height Recommendations for M301 and M302
Therefore, transfer surfaces that are adjustable to a 25 inches maximum during patient transfer accommodate most patients who use mobility devices....
-
5.1.3 Transfer Surface Low Height Options for M301 and M302
5.1.3 Transfer Surface Low Height Options for M301 and M302 The Committee did not agree on a final low height recommendation....
-
5.1.4 Transfer Surface Height Measurement for M301 and M302
5.1.4 Transfer Surface Height Measurement for M301 and M302 The Committee recommends capturing the transfer surface height at the highest point of the seat in an uncompressed state, inclusive...