5.2.1.3 Transfer Surface Size for Stretchers. The transfer surface should be 28 inches wide minimum and 17 inches deep minimum located on both long sides of stretchers....
Search Results "Stretcher"
-
5.2.1.3 Transfer Surface Size for Stretchers
-
5.3.2.1 Transfer Sides for Stretchers
5.3.2.1 Transfer Sides for Stretchers. The transfer surface shall be located to provide the ability to transfer from a mobility device onto both long sides of the surface....
-
5.4.1.1 Transfer Support Location for Stretchers
5.4.1.1 Transfer Support Location for Stretchers. The transfer surface for stretchers should be oriented along the long side of the surface....
-
5.4.2.1 Transfer Support Length for Stretchers
5.4.2.1 Transfer Support Length for Stretchers. Transfer supports on stretchers should be a minimum of 15 inches long....
-
5.4.4.1 Transfer Support Distance from Transfer Surface for Stretchers
5.4.4.1 Transfer Support Distance from Transfer Surface for Stretchers....
-
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers. The base of the equipment must provide a clearance of 39 inches wide minimum....
-
(1) Transfer Surface Location for Diagnostic Equipment Used in the Supine, Prone, or Side-lying Position
One commenter elaborated that stretchers and hospital beds are always entered from one or the other long side of the bed, not the foot end, due to obstructions at the head and foot ends...
-
c. Transfer Surface Size for Diagnostic Equipment Used by Patients in the Supine, Prone, or Side-lying Position
Committee differentiated between equipment whose transfer surface was located on the end of the equipment with transfer sides on one short side and one long side of adjoining sides and stretchers...
-
M301.4.1 Clearance in Base (Section-by-Section Analysis)
These three commenters also raised concerns that this provision was in conflict with the prevailing standard used by manufacturers for medical beds and stretchers, IEC 60601-2-52, which...
-
M305.2.1 Location (Section-by-Section Analysis)
This commenter noted that an adjacent transfer support would be more effective, but would conflict with the provider expectations of bed and stretcher side rails....
-
M305.2.2 Length (Section-by-Section Analysis)
For stretchers, the MDE Advisory Committee also recommended 15 inches in length stating that this would provide continuous support for patients and still accommodate the articulation that...
-
M301.4.1 Clearance in Base
-
Clear Floor and Turning Space Inside Examination Rooms
When a portable patient lift or stretcher is to be used, additional clear floor space will be needed to maneuver the lift or stretcher....
-
Additional Transfer Techniques: Use of Stretchers and Gurneys
from the stretcher to the table or surface....
-
Radiologic Equipment
Assisted Transfer to Densitometer Utilizing a Stretcher 1. Set stretcher height to be level with table surface for lateral transfer. 2....
-
4.2 Stretchers
4.2 Stretchers Stretchers are ubiquitous throughout inpatient, outpatient surgery, and other health care settings....
-
5.2.1.3 Transfer Surface Size Recommendations for Stretchers (M301)
5.2.1.3 Transfer Surface Size Recommendations for Stretchers (M301) The Committee recommends a minimum transfer surface size of 28 inches wide by 17 inches deep for stretchers located...
-
5.3.2.1 Transfer Sides Recommendation for Stretchers
5.3.2.1 Transfer Sides Recommendation for Stretchers On stretchers, the transfer surface is oriented along the long dimension of the surface on which patients are in the supine, prone...
-
5.4.1.1 Transfer Support Location Recommendation for Stretchers
5.4.1.1 Transfer Support Location Recommendation for Stretchers For Stretchers, the transfer surface is oriented along the long side of the surface....
-
5.4.2.1 Transfer Support Length Recommendation for Stretchers
5.4.2.1 Transfer Support Length Recommendation for Stretchers On Stretchers, the transfer surface is oriented along the long dimension of the surface on which patients are positioned...
-
5.4.4 Transfer Support Distance from Transfer Surface Recommendations for M301 and M302
5.4.4 Transfer Support Distance from Transfer Surface Recommendations for M301 and M302 Description: This minimum distance ensures that the transfer support is within reach of the...
-
5.4.4.1 Transfer Support Distance from Transfer Surface Recommendation for Stretchers
5.4.4.1 Transfer Support Distance from Transfer Surface Recommendation for Stretchers The Committee recommends locating the transfer support along the long side of the of the transfer...
-
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers (M301)
Rationale for the recommendation The recommendation for stretchers is for a smaller clearance width than the 44-inch minimum for other equipment....
-
Medical Diagnostic Equipment Accessibility Standards Sub Committee Recommendations Stretcher Final Report
Medical Diagnostic Equipment Accessibility Standards Sub Committee Recommendations Stretcher Final Report May 31, 2013 The Subcommittee’s technical recommendations for the stretchers...