M305 provides the technical requirements for supports on medical diagnostic equipment....
Search Results "Horizontal Standing Support"
Commonly Searched Documents
-
E. M305 Supports
-
M305.3.1 Horizontal Position
Where the support is horizontal, the top of the gripping surface shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface....
-
M305.3.2.1 Horizontal Position
The height of the top of the gripping surface on horizontal standing supports shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface....
-
M305.3.1.1 Horizontal Position
The length of gripping surfaces on horizontal standing supports shall be 4 inches (100 mm) minimum except for diagnostic equipment containing a wheelchair space which shall comply with M305.3.1.2...
-
Standing Supports (M304.3 and M305.3)
M305.3 also provides technical criteria for standing supports in horizontal and vertical positions....
-
M305.3.1 Length
The length of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve....
-
M304.2.2 Standing Supports
Standing supports shall be provided on two sides of the standing surface and shall comply with M305.3....
-
M305.3.1 Length (Section-by-Section Analysis)
In the MDE NPRM, the Access Board proposed a gripping surface length of four inches minimum for horizontal standing supports. No public comments were submitted on this requirement....
-
M305.3.1.2 Diagnostic Equipment Containing a Wheelchair Space
On diagnostic equipment containing wheelchair spaces with one entry that also serves as the exit, the length of the gripping surface of horizontal standing supports shall be equal to or...
-
Horizontal Dispersion: ADA Standard Section 221.2.3.1
Section 221.2.3.1 covers scoping requirements in the current ADA Standards for horizontal Dispersion in assembly areas.
-
Standing Supports (M304.3 and M305.3)
M305.3 Proposed Provision: Also provides technical criteria for standing supports in horizontal and vertical positions....
-
M304.3 Standing Supports
Standing supports shall be provided on each side of the standing surface and shall comply with M305.3....
-
M304.2.2 Standing Supports (Section-by-Section Analysis)
requirements for standing supports in proposed M305.3....
-
D. M304 Diagnostic Equipment Used by Patients in a Standing Position
M304 provides the technical requirements for diagnostic equipment used by patients in a standing position....
-
5.9.1 Standing Supports In Reference To Mammography Equipment
5.9.1 Standing Supports In Reference To Mammography Equipment. The requirement for standing supports should be removed for mammography equipment....
-
M305.3 Standing Supports
Standing supports shall provide continuous support throughout use of the diagnostic equipment and shall comply with M305.3....
-
M305.3 Standing Supports
Standing supports shall provide continuous support throughout use of the diagnostic equipment and shall comply with M305.3....
-
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position
The proposed provision requires standing supports be provided on each side of the standing surface....
-
M305.3 Standing Supports (Section-by-Section Analysis)
This provision has been reorganized in the final rule into requirements for length and height, as opposed to vertical and horizontal....
-
5.9.1 Standing Supports Recommendation In Reference To Mammography Equipment
5.9.1 Standing Supports Recommendation In Reference To Mammography Equipment The Committee recommends removing the requirement for standing supports on mammography equipment....
-
1. Standing Supports
The proposed rule included a provision requiring standing supports on each side of the standing surface and compliance with the technical requirements for standing supports in proposed M305.3...
-
5.9 Standing Support for Equipment Used by Patients in a Standing Position
5.9 Standing Support for Equipment Used by Patients in a Standing Position....
-
Back Support: ADA Standard Section 903.4
The bench shall provide for back support or shall be affixed to a wall....
-
M305 Supports
M305 provides the technical criteria for transfer supports and standing supports....