Transfer supports, leg supports, and reclining surfaces shall comply with M301.3.
Search Results "Adjustable Height Exam Table"
-
M301.3 Supports
-
M301.3 Supports
Transfer supports, stirrups, and reclining surfaces shall comply with M301.3.
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2.
-
M301.3.1 Transfer Supports
Transfer surfaces required by M301.2 shall provide transfer supports and shall comply with M305.2.
-
M301.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and shall comply with M305.4.
-
M301.3.2 Stirrups
Where stirrups are provided, they shall provide a method of supporting, positioning, and securing the patient’s legs.
-
M301.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided. Where the incline of the back support can be modified while in use, head and...
-
M301.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided and shall comply with M305.5.
-
M301.4 Lift Compatibility
Diagnostic equipment shall be usable with a patient lift and shall comply with M301.4.1 or M301.4.2.
-
M301.4 Lift Compatibility
Diagnostic equipment shall be usable with portable patient lifts and, when in use with such lifts, shall comply with M301.4.1 or M301.4.2. EXCEPTION: Where fixed overhead patient...
-
M301.4.1 Clearance in Base
The base of the equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum...
-
M301.4.1 Clearance in Base
The base of diagnostic equipment shall provide a clearance 39 inches (990 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep...
-
M301.4.2 Clearance Around Base
The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...
-
M301.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...
-
Clear Floor and Turning Space Inside Examination Rooms
The exam table must have sufficient clear floor space next to it so that an individual using a wheelchair can approach the side of the table for transfer onto it....
-
3.2 Getting on the Table
The height of exam tables was the most commonly reported complaint.4 A survey of approximately 400 Californians with disabilities found exam tables were inaccessible to 69% of wheelchair...
-
M302.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M302.2.2 Sunset
The low transfer position height, Item A of M302.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register]. [sic]...
-
Features of an Accessible Examination Room
A clear floor space, 30” X 48” minimum, adjacent to the exam table and adjoining accessible route make it possible to do a side transfer. 2....
-
M305 Supports
-
M305 Supports
-
M305.1 General
Supports shall comply with M305.
-
M305.1 General
Supports shall comply with M305, as applicable.
-
M305.2 Transfer Supports
Transfer supports shall comply with M305.2.